Features
- Input/Output optical isolation.
- Current pedestal output.
- Adjustable high and low output levels.
- Force high or low selection switch.
- Loop powered.
- Supplied with comprehensive connection lead and adaptor set.
- Detailed application information.
|
|
X101 Pulse Level Converter
A loop powered pulse converter with input/output optical isolation. The output is current pedestal with adjustable high and low levels with switch to force high or low level.
Typical Applications
- Turbine meter pre-amplifier simulation.
- Gas and Liquid Densitometer simulation.
- Coriolis meter simulation.
Allows normal signal generators to perform proper functional testing of panel conditioning and instrument inputs with minimal disturbance to panel wiring.
Specification
- Input: ±15 Volt maximum.
- Output: 30 Volt dc maximum; Off, Low level 1-12mA, High level 2-20mA (+low level).
- Frequency: 0 - 5 kHz.
- Isolation: >10 MΩ Input to Output.
- Size: 93 x 38 x 30 mm (overall including connectors and switch).
|